GE Aespire S5 Anesthesia Machine with 7900 Smartvent_1

GE Aespire S5 Anesthesia Machine with 7900 Smartvent



Clinician-inspired solutions help improve patient care

Features

  • Enhanced monitor integration capabilities with the GE / Datex Ohmeda Anesthesia Monitor Series.
  • Lightweight and compact for easy maneuverability
  • Optional integrated auxiliary O2 flowmeter and suction control

Advanced Breathing System (ABS™)

  • One step bag/vent switch turns the ventilator on/off
  • Minimal number of parts and tube connections greatly reduces the potential for leaks and misconnects
  • Ease of disassembly (no tools)
  • Fully autoclavable and latex-free

7900 SmartVent Ventilator

  • Volume and Pressure modes with electronic PEEP
  • Synced intermittent mandatory ventilation ( SIMV )
  • Direct access to ventilator parameter settings
  • PSVPro ( Pressure support with Apnea backup mode. )
  • Smart alarms direct user to specific problems and affected parameters
  • Ideal for spontaneously breathing patients

Improved low flow/reduced life cycle costs

  • Only one scheduled maintenance check per year
  • Fresh gas flow compensation – automatically (available with tidal volume compensation option)
  • Minimum O2 flow of 50 mL
  • Dual air flow tubes standard for higher resolution of low flows

SPECIFICATIONS

Physical Weight & Dimensions

  • Height: 134.5 cm/52.9 in
  • Width: 72 cm/28.3 in
  • Depth: 73 cm/28.7 in
  • Weight: Approximately 108 kg/238 lb

Top Shelf

  • Weight limit: 34 kg/75 lb
  • Width: 66 cm/26 in
  • Depth: 40 cm/15.75 in

Work Surface

  • Height: 81.7 cm/32.2 in
  • Size: 2160 cm-2/334 in-2

DIN Rail

Side of machine: 34.5 cm/13.6 in

Drawers (internal dimensions)

  • Height: 17.5 cm/6.9 in
  • Width: 33 cm/13 in
  • Depth: 26.5 cm/10.4 in

Absorber Bag Arm (optional)

  • Arm length: 30.5 cm/12 in
  • Bag arm height (adjustable): 87 cm (34.3 i) to 104 cm (40.9 in)

Casters

  • Diameter: 12.5 cm/5 in
  • Brakes: Individual locking front casters

Electrical Specifications

Current Leakage

  • 100/120 V: < 300μA
  • Power
  • Power input: 100-120 Vac, 50/60 Hz
  • Power cord: Length: 5 m/16.4 ft
  • Rating: 15A @ 120 Vac

Inlet/outlet modules (120 V)

  • System circuit breakers: 15A
  • Outlets (optional): 4 outlets on back, 3-2A, 1-3A individual breakers, optional isolation transformer

Pneumatic Specifications

Auxiliary Common Gas Outlet

Connector: ISO 22 mm OD and 15 mm ID

Gas Supply

  • Pipeline input range: 240 kPa to 600 kPa/35 psig to 88 psig
  • Pipeline connections: DISS-male, DISS-female, DIN 13252, AS4059, S90-116 or NIST (ISO 5359). All fittings available for O2, N2O, and Air, and contain pipeline filter and check valve.
  • Cylinder input: Pin indexed in accordance with CGA-V-1 or DIN (nut and gland); contains input filter and check valve
    Note: Maximum 3 cylinders; two inboard mounted, one outboard mounted.
  • Primary regulator diaphragm minimum burst pressure: 2758 kPa/400 psig
  • Primary regulator nominal output: < 338 kPa/49 psig Pin indexed cylinder connections
  • < 407 kPa/59 psig DIN cylinder connections

Environmental Specifications

System Operation

  • Temperature: 10° to 40°C/50° to 104°F
  • Humidity: 15 to 95% relative humidity (non-condensing) per IEC 68-2-3
  • Altitude: -440 to 3565 m/500 to 800 mmHg

System Storage

  • Temperature: -15° to 50°C/-5° to 122°F
  • Humidity: 10 to 95% relative humidity (including condensing) per IEC 68-2-3
  • Altitude: -440 to 3565 m/500 to 800 mmHg
  • Oxygen cell storage: -15° to 50°C/5° to 122°F; 10 to 95% relative humidity; 500 to 800 mmHg

Electromagnetic Compatibility

  • Immunity: Complies with all requirements of EN 60601-1-2
  • Emissions: CISPR 11 group 1 class B
  • Approvals: UL 2601-1; CSA C22.2 #601.1; EN/IEC 60601-1; CE 0197